The 107TD47 (B92016003) is an insulation monitoring device with load current and temperature monitoring, designed for IT systems in medical locations according to DIN VDE 0100-710. This A-ISOMETER operates on an AC supply voltage of AC 0-264V (US: AC 230V). It features insulation monitoring with an alarm range of R-ALARM 50-500 kOhm, and load monitoring with an alarm range of I-ALARM 5-50 A (functioning exclusively in connection with STW2). Additionally, it provides temperature monitoring via a DIN44081 PTC Thermistor. The unit is equipped with an illuminated LCD display and a BMS bus. Note that this part is obsolete and has been replaced by B72075301.

More Bender part numbers

See all Bender products

Frequently Asked Questions(FAQ):


What is the price of the 107TD47 (B92016003) obsolete/replaced by B72075301?

For customized pricing and availability, please fill out ourcontact formor give us a call. We are dedicated to providing solutions that align with your specific requirements.


While we are not the official distributor or seller ofBenderin the United States, we source our products from reliable channels and offer you originalBenderequipment at competitive prices.

How can I access 107TD47 (B92016003) obsolete/replaced by B72075301 catalogs, user manuals, technical documents, and PDFs?


You can access catalogs, user manuals, and technical documents for107TD47 (B92016003) obsolete/replaced by B72075301on our website. Visit our resources section or contact us directly for specific documents.

Do you provide technical support for 107TD47 (B92016003) obsolete/replaced by B72075301?


Do you provide technical support for 107TD47 (B92016003) obsolete/replaced by B72075301?

While we do not offer direct technical support, we can assist you in contacting the manufacturer to ensure you receive the support you need.

Characteristics

Manufacturer Bender
Country
Part number 107TD47 (B92016003) obsolete/replaced by B72075301
Inner Code 26802

Technical specification

Product type Insulation monitoring device
Supply voltage AC 0-264V (US: AC 230V)
Standard DIN VDE 0100-710
Application Medical spaces
Insulation monitoring R-Alarm 50-500 kOhm
Load monitoring I-Alarm 5-50 A (only in connection with STW2)
Temperature monitoring DIN44081 PTC Thermistor
Display Illuminated LCD display
Interface BMS bus
Replacement part B72075301