The 107TD47 (B92016003) is an insulation monitoring device with load current and temperature monitoring, designed for IT systems in medical locations according to DIN VDE 0100-710. This A-ISOMETER operates on an AC supply voltage of AC 0-264V (US: AC 230V). It features insulation monitoring with an alarm range of R-ALARM 50-500 kOhm, and load monitoring with an alarm range of I-ALARM 5-50 A (functioning exclusively in connection with STW2). Additionally, it provides temperature monitoring via a DIN44081 PTC Thermistor. The unit is equipped with an illuminated LCD display and a BMS bus. Note that this part is obsolete and has been replaced by B72075301.
For customized pricing and availability, please fill out ourcontact formor give us a call. We are dedicated to providing solutions that align with your specific requirements.
How can I access 107TD47 (B92016003) obsolete/replaced by B72075301 catalogs, user manuals, technical documents, and PDFs?
Do you provide technical support for 107TD47 (B92016003) obsolete/replaced by B72075301?
While we do not offer direct technical support, we can assist you in contacting the manufacturer to ensure you receive the support you need.
| Manufacturer | Bender |
|---|---|
| Country | |
| Part number | 107TD47 (B92016003) obsolete/replaced by B72075301 |
| Inner Code | 26802 |
| Product type | Insulation monitoring device |
|---|---|
| Supply voltage | AC 0-264V (US: AC 230V) |
| Standard | DIN VDE 0100-710 |
| Application | Medical spaces |
| Insulation monitoring R-Alarm | 50-500 kOhm |
| Load monitoring I-Alarm | 5-50 A (only in connection with STW2) |
| Temperature monitoring | DIN44081 PTC Thermistor |
| Display | Illuminated LCD display |
| Interface | BMS bus |
| Replacement part | B72075301 |