10112334 19 1603 1127 by Msa
Get a quote
The MSA part number 10112334 19 1603 1127 is an explosion-protected PrimaX P HART-EX transmitter housed in a robust powder-coated aluminum junction enclosure designed for monitoring hydrogen sulfide (H2S) in air. Operating on an electrochemical measuring cell principle, this unit is factory-calibrated for H2S with a measuring range of 0 to 50 ppm, alongside adjustable ranges of 0 to 10 ppm, 0 to 20 ppm, and 0 to 100 ppm configurable via the control panel. It features a built-in LCD display, 3-wire wiring, outputs for 4 to 20 mA, alarm relays, and fault relays, plus an intrinsically safe HART interface utilizing the digital HART protocol for evaluation and configuration. The device includes preset alarms and operates on a typical power consumption of 24 VDC at 3 W. Certified for hazardous locations under ATEX and IECEx standards, the unit measures 220 x 162 x 100 mm, weighs 1.6 kg, and comes supplied with a nickel-plated brass M25x1.5 mm Ex d cable gland, splash guard, and calibration cap.
More Msa part numbers
Frequently Asked Questions(FAQ):
Where can I buy 10112334 19 1603 1127 by Msa in the USA?
You can purchase 10112334 19 1603 1127 by Msa from PAOParts LLC. We serve customers across the USA, offering competitive prices and reliable service.
What is the price of the 10112334 19 1603 1127?
Price and availability are quoted per request. Send the part number and the quantity you need through our quote request form and you will receive a written offer by e-mail.
Who is the official distributor or seller of Msa in the United States?
While we are not the official distributor or seller of Msa in the United States, we source our products from reliable channels and offer you original Msa equipment at competitive prices.
How can I access 10112334 19 1603 1127 catalogs, user manuals, technical documents, and PDFs?
Ask for them in your quote request. We attach the datasheet, catalogue page or manual we hold for 10112334 19 1603 1127 to our written reply.
Do you provide technical support for 10112334 19 1603 1127?
While we do not offer direct technical support, we can assist you in contacting the manufacturer to ensure you receive the support you need.
Characteristics
| Manufacturer | Msa |
|---|---|
| Part number | 10112334 19 1603 1127 |
| Inner Code | 589095 |
Technical specification
| Product type | Explosion-Protected HART-EX Transmitter |
|---|---|
| Manufacturer | MSA |
| Part Number | 10112334 19 1603 1127 |
| Model | PrimaX P, M25, H+R, H2S 50 |
| Enclosure Material | Powder-coated aluminum |
| ATEX Protection Class Gas | II 2G Ex d ia [ia] IIC T4 Gb |
| ATEX Protection Class Dust | II 2D Ex tb ia [ia] IIIC T130°C Db IP67 |
| Ambient Temperature Range ATEX | -40°C to +55°C (Ox-Tox-Sensor +70°C) |
| Type Examination Certificate Number | BVS 10 ATEX E 009 X |
| IECEx Certificate Number | IECEx BVS 10.0063X N1 |
| Enclosure Protection Class | IP 66 |
| Measuring Principle | Electrochemical measuring cell |
| Calibrated Gas | Hydrogen sulfide (H2S) in air |
| Measuring Range | 0...50 ppm |
| Transmitter Alarm | 30 % of measuring range |
| Evaluation Unit Alarm 1 | 5 ppm |
| Evaluation Unit Alarm 2 | 10 ppm |
| Configurable Measuring Ranges | 0...10 ppm, 0...20 ppm, 0...100 ppm |
| Display | LCD |
| Outputs | 4...20 mA, Alarm relay, Fault relay |
| Interface | Intrinsically safe HART interface, digital HART protocol |
| Wiring | 3-wire |
| Permissible Ambient Temperature Range | -40...+50°C |
| Relative Humidity | 15...90% non-condensing |
| Pressure Range | 80...120 kPa |
| Power Consumption | 24 VDC / 3 W typical |
| Dimensions (H x W x D) | 220 x 162 x 100 mm |
| Weight | 1.6 kg |
| Included Items | Cable gland M25x1.5 mm Ex d 9-16 mm Brass nickel-plated, Splash protection, Calibration cap |